Application
ROHS,toy safety,electronics,ceramics and plastics,leather jewelry,environmental protection,food safety,plating thickness,metal alloy and so on.
Technical specification
l Element content:2 PPM - 99.99%;
l Measurable element: element S (sulfur) - element U (uranium),total 75 elements.
l Measurement time: 100-300 Seconds
l Energy Resolution: 149 ± 5 ev
l X-ray tube maximum output current: 1mA;
l High voltage maximum output power: 50w(Voltage:5-50Kv Current:50-1000uA)
l Sample chamber size: 610mm*320mm*100mm
l Volume: 700mm*430mm*400mm
l Weight: 68kgs
Main Feature
- Si-pin Electric Cooling Semiconductor Detectors: (The Newest Detector Type) Si-pin Electric Cooling semiconductor detector;Resolution: 149±5 Electron Volts Amplifier Circuit Modules:Detect the sample feature X-ray;Enlarge the collected Information.
- X-ray Excitation Device: Filament Current Maximum Output: 1mA; Belongs to the semi-loss components,Lifetime use more than 20,000 hours.
- High Voltage Transmit Device :
- Maximum Voltage Output: 50kV.
- Minimum 5kv Controllable Adjustment.
- Over voltage Protection by itself.
- Muti-channel Analyzer:
- Convert signal into a digital signal,and send the results to the host computer Software.
- Maximum Channels Number:4096.
- Including the signal enhancement processing.
- Light Path Filtering Module
- Reduce the interference of the X-ray light send,ensure the accurate detector signal data.
- Integration the collimator and the filter.
- Collimator Automatically Switch Module
- Up to six options, caliber are 8-1#, 8-2#, 8-3#, 3# ,1#, 1-2v#.
- Filter Automatically Switch Module
- Free selection and switching of the five kinds filter.
- Free Combination Module between Collimator and Filter
- Up to dozens free combination between collimator and filter.
- Automatically Curve Selected Module
- Select the working curve automatically,without manually operation.Avoid human error,perfect combine the automation with the intelligent operation,make it more user-friendly, and more convenient.
Working Conditions
Working temperature:15-30℃
Relative humidity: 40%~50%
Power Source: AC :220V ±5V
Sample configuration
- Standard samples be used to making working curve
- Sample Chamber:Open Type Large style sample chamber
- Standard Sample:EU Standard Sample EC680K
- Pure silver samples
Warranty and After-sale Service
- Free Training to customers.
- Free training in user’s location about installation,debug,test,training and technical services.
- We provide one year warranty against normal daily use problems or material defects,but the term is there are no other person disassemble action besides our own engineers.
- Lifelong Maintenance.
- Provide Free Software Upgrades;
- Provide the most effective technical services,and respond customers within 24hours.