Instrument Introduction
Features
● EDX8300H is equipped with vacuum to better cope with environmental protection directives such as RoHS and halogen-free. It is equipped with a vacuum test system, which increases the scope of element testing. The detector is equipped with a test system for SDD, especially for improving Na, Mg, Al, Si. The detection limit of light elements is improved, the accuracy is improved, and the repeatability of other elements such as Cd/Pb/Cr/Hg/Br/Cl is significantly improved.
● The overall structural design, the instrument is beautiful and generous.
● Using the latest Si-pin detector from USA, it is electrically cooled rather than liquid nitrogen cooled, with small size, accurate data analysis and low maintenance cost.
● Adopt self-developed SES signal processing system to effectively improve the peak-to-back ratio and make the measurement more accurate.
● One-button automatic test, the use is simpler, more convenient, and the interface is more user-friendly.
● Seven kinds of optical path correction collimation systems, automatically switch according to different samples.
● Multiple radiation protection design, the radiation protection level is the highest level of similar products.
● Optimize the integrated heat dissipation design to make the heat dissipation performance of the whole machine more improved and ensure the safe operation of the X-ray source.
● Unique movement temperature monitoring technology ensures the safe and reliable operation of the radiation source, effectively extending its service life and reducing the cost of use.
● Multiple instrument parts protection system, and can be monitored by software to make the instrument work more stable and safer.
● Special vacuum test software, whether vacuum can be freely selected, vacuum degree real-time display, standard window design, friendly interface, easy operation, all control is between the mouse.
● This machine adopts USB2.0 interface to ensure accurate and high-speed and efficient data transmission.
● Unique automatic vacuum system completely shields the influence of air on light element test, greatly improves the detection limit of light elements, and also expands the detection range.
● Integrated vacuum chamber design, effectively ensuring the vacuum level in accordance with the test requirements, and improving the maintainability of the equipment.
● Vacuum test and non-vacuum test The conversion is simple and convenient for customers. At the same time, the vacuum is released when the cover is opened, which is safe, simple and practical.
Working Condition
Working temperature: 15-30 ° C
Relative humidity: 40% to 70%
Power: AC: 220V ± 5V
Technique Specifications
● Elemental analysis ranges from sodium (Na) to uranium (U)
● Elemental content analysis range is from 1PPm to 99.99%
● Measurement time: 100-300 seconds (adjustable)
●The harmful elements (Cd/Pb/Cr/Hg/Br/Cl) specified in the RoHS Directive have detection limits of up to 1PPM
● The energy resolution is 149±5eV
● Temperature adaptation range is 15°C to 30°C
● Power supply: AC 220V±5V (recommended configuration AC purification power supply)
Instrument Configuration
SDD Electric Refrigeration Detector
● SDD electric refrigeration detector; resolution: 129±5e eV
● Amplifying circuit module: detecting X-rays of sample features; further amplifying the information collected by the detection.
X-ray Excitation Device
● Filament current maximum output: 1mA;
● It is a semi-loss type component with a service life of more than 5000 hours.
High Pressure Launcher
● Maximum voltage output: 50kV
● Minimum 5kv controllable adjustment
● Built-in voltage overload protection
Integrated Vacuum System
● Vacuum pump system with excellent performance such as low vibration, low noise, self-protection and fast pumping speed;
● Geometric pumping speed: 60 L/min (50Hz)
● Ultimate pressure: 6.7×10-2 Pa
Multichannel Analyzer
● Converting the collected analog signal into a digital signal, and providing the processing result to the host computer software
● The maximum number of tracks: 4096
● Includes signal enhancement processing
Optical Path Filter Module
● Reduce the interference during the transmission of the X-ray optical path to ensure that the detector receives the signal accurately
● Integrating the collimator with the filter
Test Software MeaRohs
Function
● Specifically for the five elements involved in the EU RoHS Directive Cd, Pb, Hg, Br, Cr and halogen element Cl test, measuring time is 100-300 seconds (adjustable).
● The operation interface is simple and intuitive, easy to use, no need for professional operation.
Features
● Automatically select the appropriate calibration curve for easier and more accurate measurement.
● Chinese and English interface automatically switch, and has third-party language customization.
● Automatic calibration instrument
● Self-contained sample material qualitative analysis to prevent manual users from selecting wrong curves
● Print in multiple report formats
● Multiple spectra can be displayed simultaneously
● Real-time display of instrument test parameters and vacuum, all control at a glance
● The unique movement temperature monitoring technology ensures the safe and reliable operation of the X-ray source, effectively extending its service life and reducing the cost of use.
Sample Configuration
Sample Chamber
● Open large sample chamber: 610mm × 320mm × 100mm
● Semi-closed sample chamber (when vacuuming): Φ100mm×h75mm
Standard Samples
● EU standard EC680K, EC681K
● Silver calibration sample
Standard Accessories
Low noise fast pumping speed vacuum pump
● Small size, light weight, simple structure, easy maintenance and repair
●The pumping speed is fast, 20S can reach the vacuum test requirements.
● The ultimate pressure can reach 6.7×10-2Pa
● Single-phase AC220V power supply, no external three-phase power required
;